Huakeyi HK-1303 Conductivity Analyzer

HK-1303 Conductivity Analyzer

HK-1303 Conductivity Analyzer

480×272 color lattice LCD display in English and Chinese. Intelligence programmable, setup for output current, alarm and electrode optional, and customize measurement interface according to needs. 4 electrodes with 4 cell constants, each electrode corresponds to 4 ranges. Full Cast aluminum metal case eliminates the electromagnetic interference. Manual and automatic temperature compensation mode can avoids the effect of water temperature. PT1000 high performance platinum resistance temperature device makes measurement more reliable and accurate.

Overview

  • Supplies category : Water Analysis
  • Material code : HK-1303
  • Manufacturers : Beijing Huakeyi Power Plant Instrument Research Institute

Overview

  • Supplies category : Water Analysis
  • Material code : HK-1303
  • Manufacturers : Beijing Huakeyi Power Plant Instrument Research Institute

Specifications
Features
Specifications

Display

480 × 272 color lattice LCD display in English or Chinese

Measurement Range

K=0.01 (0.000~0.200)μS/cm,(0.000~2.000)μS/cm,(0.000~20.00)μS/cm, (0.000~200.0)μS/cm 
K=0.10 (0.000~2.000)μS/cm,(0.000~20.00)μS/cm,(0.000~200.0)μS/cm, (0.000~2000)μS/cm
K=1.00 (0.000~20.00)μS/cm,(0.000~200.0)μS/cm,(0.000~2000)μS/cm, (0.000~20.00)mS/cm
K=10 (0.000~200.0)μS/cm,(0.000~2000)μS/cm,(0.000~20.00)mS/cm, (0.000~200.0)mS/cm

Display Resolution

0.001μS/cm

Analysis Error

±1.5%FS

Response Time (T90)

30s at 25℃ ambient temperature

Temperature Sensor

Pt1000

Sample Condition

Temperature: 5 to 50oC Flow: (10-40)L/h Solid contents in sample: ≤5mm (no gel and grease)

Temperature Range

(0.0 ~ 99.9)℃

Temperature Accuracy

±0.6℃

Temperature Resolution

0.1℃

Temperature Compensation

(0.1 ~ 60.0)℃ (auto or manual)

Comp. Temp. Coefficient

(0.0~9.99)%/℃

Current Output

(0~10)mA, (0~20)mA or (4~20)mA (two isolated outputs)

Current Output Error

±1%FS

Max load resistance

800Ω(20mA max.)

Power Supply

(100~240)VAC, frequency(50~60)Hz

Power Consumption

≤15W

Insulation

Power to ground 1,500V, no electric arcing in 1 min

Alarm output

two relay outputs programmable, 250VAC/5A or 30VDC/5A

Storage Temperature

(-20 ~ 55)℃

Transmitter Dimension (W×H×D)

(146×146×154)mm

Cutout dimension

(139×139)mm

Weight

2 kg

 

Features
  • 480×272 color lattice LCD display in English and Chinese.

  • Intelligence programmable, setup for output current, alarm and electrode optional, and customize measurement interface according to needs.

  • 4 electrodes with 4 cell constants, each electrode corresponds to 4 ranges.

  • Full Cast aluminum metal case eliminates the electromagnetic interference.

  • Manual and automatic temperature compensation mode can avoids the effect of water temperature.

  • PT1000 high performance platinum resistance temperature device makes measurement more reliable and accurate.

For further explanation, please contact our marketing support.

Let’s work together with

Hidro Dinamika Kencana

water treatment specialist in Indonesia

Engineering  |  Procurement  |  Construction

Let’s work together with

Hidro Dinamika Kencana

water treatment specialist in Indonesia

Engineering  |  Procurement  |  Construction